Objectives: The goal of this study was to describe early and midterm outcomes of extremely premature newborns (EPNs) who underwent transcatheter echocardiographically guided patent ductus arteriosus (PDA) closure.
Background: Surgical ligation of PDA in EPNs confers significant risk for procedural morbidity and adverse long-term outcomes.
Methods: The Amplatzer Vascular Plug II was used in all cases. Post-ligation syndrome was defined using previously published parameters. Patients were followed at pre-specified intervals, and prospectively collected data were reviewed.
Results: Transcatheter closure was attempted in 24 EPNs (mean procedural age 30 days [range 5 to 80 days], mean procedural weight 1,249 g [range 755 to 2,380 g]) and was successful in 88%. The 3 procedural failures were related to the development of left pulmonary artery (LPA) stenosis caused by the device, and all devices were removed uneventfully. Complications included 2 instances of device malposition, resolved with device repositioning, and 1 instance of LPA stenosis, requiring an LPA stent. There were no procedural deaths, cases of post-ligation syndrome, residual PDA, or device embolization. Survival to discharge was 96% (23 of 24), with a single late death unrelated to the procedure. After a median follow-up period of 11.1 months, all patients were alive and well, with no residual PDA or evidence of LPA or aortic coarctation.
Conclusions: This newly described technique can be performed safely with a high success rate and minimal procedural morbidity in EPNs. Early and midterm follow-up is encouraging. Future efforts should be directed toward developing specific devices for this unique application.
Keywords: device occlusion; patent ductus arteriosus; premature newborn.
Copyright © 2016 American College of Cardiology Foundation. Published by Elsevier Inc. All rights reserved.