Transcatheter Closure of Patent Ductus Arteriosus in Extremely Premature Newborns: Early Results and Midterm Follow-Up

JACC Cardiovasc Interv. 2016 Dec 12;9(23):2429-2437. doi: 10.1016/j.jcin.2016.09.019.

Abstract

Objectives: The goal of this study was to describe early and midterm outcomes of extremely premature newborns (EPNs) who underwent transcatheter echocardiographically guided patent ductus arteriosus (PDA) closure.

Background: Surgical ligation of PDA in EPNs confers significant risk for procedural morbidity and adverse long-term outcomes.

Methods: The Amplatzer Vascular Plug II was used in all cases. Post-ligation syndrome was defined using previously published parameters. Patients were followed at pre-specified intervals, and prospectively collected data were reviewed.

Results: Transcatheter closure was attempted in 24 EPNs (mean procedural age 30 days [range 5 to 80 days], mean procedural weight 1,249 g [range 755 to 2,380 g]) and was successful in 88%. The 3 procedural failures were related to the development of left pulmonary artery (LPA) stenosis caused by the device, and all devices were removed uneventfully. Complications included 2 instances of device malposition, resolved with device repositioning, and 1 instance of LPA stenosis, requiring an LPA stent. There were no procedural deaths, cases of post-ligation syndrome, residual PDA, or device embolization. Survival to discharge was 96% (23 of 24), with a single late death unrelated to the procedure. After a median follow-up period of 11.1 months, all patients were alive and well, with no residual PDA or evidence of LPA or aortic coarctation.

Conclusions: This newly described technique can be performed safely with a high success rate and minimal procedural morbidity in EPNs. Early and midterm follow-up is encouraging. Future efforts should be directed toward developing specific devices for this unique application.

Keywords: device occlusion; patent ductus arteriosus; premature newborn.

Publication types

  • Research Support, Non-U.S. Gov't

MeSH terms

  • Cardiac Catheterization* / adverse effects
  • Cardiac Catheterization* / instrumentation
  • Cardiac Catheterization* / mortality
  • Ductus Arteriosus, Patent / diagnostic imaging
  • Ductus Arteriosus, Patent / mortality
  • Ductus Arteriosus, Patent / physiopathology
  • Ductus Arteriosus, Patent / therapy*
  • Echocardiography, Doppler, Color
  • Echocardiography, Doppler, Pulsed
  • Gestational Age
  • Humans
  • Infant, Extremely Premature*
  • Infant, Newborn
  • Predictive Value of Tests
  • Retrospective Studies
  • Risk Factors
  • Time Factors
  • Treatment Outcome