Data for a pre-performance test of self-developed electronic tongue sensors

Data Brief. 2016 Nov 19:9:1090-1093. doi: 10.1016/j.dib.2016.11.041. eCollection 2016 Dec.

Abstract

This article presents data, which can be applied for a pre-performance test of self-developed electronic tongue sensors. Contained data is related to the research article "Impact of Sodium Lauryl Sulfate in oral liquids on E-Tongue Measurements" (http://dx.doi.org/10.1016/j.ijpharm.2016.10.045; (L.I. Immohr, R. Turner, M. Pein-Hackelbusch, 2016) [1]). Sensor responses were obtained from 10 subsequent measurements and four different concentrations of quinine hydrochloride by electronic tongue (TS-5000Z, Insent Inc., Atsugi-Shi, Japan) measurements. Based on the data for the pre-performance testing, which were calculated based on the fluctuation range of the sensor responses around the median, stability criteria and required preconditions cycles were defined.

Keywords: Conditioning cycles; Performance testing; e-tongue.