Reconfigurable Multiparameter Biosignal Acquisition SoC for Low Power Wearable Platform

Sensors (Basel). 2016 Nov 25;16(12):2002. doi: 10.3390/s16122002.

Abstract

A low power and low noise reconfigurable analog front-end (AFE) system on a chip (SoC) for biosignal acquisition is presented. The presented AFE can be reconfigured for use in electropotential, bioimpedance, electrochemical, and photoelectrical modes. The advanced healthcare services based on multiparameter physiological biosignals can be easily implemented with these multimodal and highly reconfigurable features of the proposed system. The reconfigurable gain and input referred noise of the core instrumentation amplifier block are 25 dB to 52 dB, and 1 μVRMS, respectively. The power consumption of the analog blocks in one readout channel is less than 52 μW. The reconfigurable capability among various modes of applications including electrocardiogram, blood glucose concentration, respiration, and photoplethysmography are shown experimentally.

Keywords: analog front-end; bioimpedance; biosignal acquisition; blood glucose concentration; electrocardiogram; photoplethysmography; reconfigurable circuit.

MeSH terms

  • Biosensing Techniques / methods*
  • Blood Glucose / analysis
  • Electrocardiography / methods
  • Humans
  • Photoplethysmography / methods

Substances

  • Blood Glucose