Determination of Colistin and Colistimethate Levels in Human Plasma and Urine by High-Performance Liquid Chromatography-Tandem Mass Spectrometry

Ther Drug Monit. 2016 Dec;38(6):796-803. doi: 10.1097/FTD.0000000000000345.

Abstract

Background: Colistin is a polypeptide antibiotic from the polymyxin E group used for the treatment of infections caused by multidrug-resistant gram-negative bacteria. The main constituents, accounting for approximately 85% of this mixture, are colistin A (polymyxin E1) and colistin B (polymyxin E2). The aim of this study was to develop and validate new and fast methods of quantification of colistin A and B and its precursors [colistin methanesulfonate sodium (CMS) A and B] by ultraperformance liquid chromatography-tandem mass spectrometry in plasma and urine with short pretreatment and run times.

Methods: Chromatography was performed on an Acquity UPLC-MS/MS system (WATERS) with a WATERS Acquity UPLC C18 column (4.6 × 150 mm, 3.5 μm particle size). The pretreatment of samples consists of precipitation and extraction into microcolumns plate and HLB 96-well plate 30 μm-30 mg (OASIS) with a Positive Pressure-96 (WATERS).

Results: Quantification was performed using a multiple reaction monitoring of the following transitions: m/z 390.9 → 385.1 for colistin A, m/z 386.2 → 101.0 for colistin B, and m/z 602.4 → 241.1 for polymyxin B1 sulfate. In plasma and urine, calibration curves were linear from 30 to 6000 ng/mL for colistin A and from 15 to 3000 ng/mL for colistin B. With an acceptable accuracy and precision, the lower limit of quantification were set at 24.0 ng/mL and 12.0 ng/mL for colistin A and B in plasma, and at 18.0 ng/mL and 9.0 ng/mL for colistin A and B in urine.

Conclusions: These LC-MS/MS methods of quantification for colistin A and B and its precursors (CMS A and B) in plasma and urine are fast, simple, specific, sensitive, accurate, precise, and reliable. Furthermore, they are linear and repeatable. These procedures were successfully applied to a pharmacokinetic study of a critically ill patient suffering from ventilator-associated pneumonia, who was treated with nebulized CMS.

MeSH terms

  • Anti-Bacterial Agents / blood
  • Anti-Bacterial Agents / urine
  • Calibration
  • Chromatography, High Pressure Liquid / methods
  • Colistin / analogs & derivatives*
  • Colistin / blood*
  • Colistin / urine*
  • Drug Monitoring / methods
  • Humans
  • Polymyxins / blood
  • Polymyxins / urine
  • Tandem Mass Spectrometry / methods

Substances

  • Anti-Bacterial Agents
  • Polymyxins
  • colistinmethanesulfonic acid
  • Colistin