Effect of Double Screw on Abutment Screw Loosening in Single-Implant Prostheses

Int J Prosthodont. 2016 Sep-Oct;29(5):445-7. doi: 10.11607/ijp.4791.

Abstract

Purpose: This study was conducted to assess the effect of using a double screw on the prevention of abutment screw loosening.

Materials and methods: Internal connected abutment with a single screw (IS), internal connected abutment with a double screw (ID), external connected abutment with a single screw (ES), and external connected abutment with a double screw (ED) groups were prepared (n = 10 in each group). After 50,000 loading cycles, postload removal torque loss (RTL) percentage was measured.

Results: Postload RTL of ID and ED were smaller than those of IS and ES (P = .000 and P = .039, respectively).

Conclusion: This study showed that a double screw was more effective in prevention of screw loosening.

Publication types

  • Comparative Study

MeSH terms

  • Bite Force
  • Dental Implant-Abutment Design / instrumentation*
  • Dental Implants, Single-Tooth*
  • Dental Prosthesis Retention / instrumentation*
  • Dental Restoration Failure*
  • Humans
  • Reproducibility of Results
  • Stress, Mechanical
  • Torque