Characterization and Validation of the LT-SYS Copper Assay on a Roche Cobas 8000 c502 Analyzer

Lab Med. 2017 Feb;48(1):10-17. doi: 10.1093/labmed/lmw049. Epub 2016 Sep 4.

Abstract

Objective: Validation of the LT-SYS quantitative in vitro copper assay on a Roche Cobas 8000 c502 analyzer and comparison with a BIOMED assay on a Roche Cobas Mira analyzer.

Methods: Imprecision and bias were quantified at different concentration levels (serum and plasma) over a 20-day period. Linearity was assessed covering a range from 4.08 µmol/L to 33.8 µmol/L. Limit of blank (LoB) and limit of detection (LoD) were established based on a total of 120 blank and low-level samples. The method comparison was based on 58 plasma samples.

Results: Within-run imprecision ranged from 0.7% to 1.2% and within-laboratory imprecision from 1.4% to 3.3%. Relative bias for the 2 serum pools with known target values was less than 2.5%. The assay did not deviate from linearity over the tested measuring range. LoB and LoD were 0.12 µmol/L and 0.23 µmol/L, respectively. The method comparison revealed an average deviation of 11.5% (2.016 µmol/L), and the linear regression fit was y = 1.464 + 0.795x.

Conclusions: The LT-SYS copper assay characterized in this study showed a fully acceptable performance with good degrees of imprecision and bias, no deviation from linearity in the relevant measuring rangem, and very low LoB and LoD.

Keywords: Morbus Wilson; Roche Cobas 8000 c502; clinical chemistry; copper; method comparison; method validation.

MeSH terms

  • Adult
  • Clinical Chemistry Tests / instrumentation
  • Clinical Chemistry Tests / methods*
  • Clinical Chemistry Tests / standards*
  • Copper / blood*
  • Female
  • Humans
  • Limit of Detection
  • Linear Models
  • Male
  • Reference Standards
  • Reproducibility of Results

Substances

  • Copper