Chemical Shift magnetization transfer magnetic resonance imaging

Magn Reson Med. 2017 Aug;78(2):656-663. doi: 10.1002/mrm.26383. Epub 2016 Aug 31.

Abstract

Purpose: The purpose of this work was to develop a chemical shift magnetization transfer (CSMT) magnetic resonance imaging (MRI) method to provide accurate magnetization transfer ratio (MTR) measurements in the presence of fat.

Methods: Numerical simulations were performed to compare MTR measurements at different echo times (TEs) for voxels with varying fat/water content. The CSMT approach was developed using water fraction estimates to correct for the impact of fat signal upon observed MTR measurements. The CSMT method was validated with oil/agarose phantom and animal studies.

Results: Simulations demonstrated that the observed MTRs vary with water fraction as well as with the TE-dependent phase difference between fat and water signals; simulations also showed that a linear relationship exists between MTR and water fraction when fat and water signals are in phase. For phantom studies, observed MTR decreased with increasing oil fraction: 42.41 ± 0.54, 38.12 ± 0.33, 32.93 ± 0.56, and 26.08 ± 0.87 for 5% to 40% oil fractions, respectively, compared to 42.63 ± 1.04 for phantom containing 4% agarose only. These offsets were readily corrected with the additional acquisition of a water fraction map.

Conclusion: Fat fraction and TE can significantly impact observed MTR measurements. The new CSMT approach offers the potential to eliminate the effects of fat upon MTR measurements. Magn Reson Med 78:656-663, 2017. © 2016 International Society for Magnetic Resonance in Medicine.

Keywords: chemical shift; fat; fat water separation; magnetization transfer.

Publication types

  • Research Support, N.I.H., Extramural

MeSH terms

  • Animals
  • Computer Simulation
  • Fats / chemistry*
  • Image Processing, Computer-Assisted / methods*
  • Magnetic Resonance Imaging / methods*
  • Phantoms, Imaging
  • Rabbits
  • Water / chemistry*