Diffusion-weighted echo planar MR imaging of the neck at 3 T using integrated shimming: comparison of MR sequence techniques for reducing artifacts caused by magnetic-field inhomogeneities

MAGMA. 2017 Feb;30(1):57-63. doi: 10.1007/s10334-016-0582-z. Epub 2016 Aug 8.

Abstract

Objective: Our objective was to compare available techniques reducing artifacts in echo planar imaging (EPI)-based diffusion-weighed magnetic resonance imaging MRI (DWI) of the neck at 3 Tesla caused by B0-field inhomogeneities.

Materials and methods: A cylindrical fat-water phantom was equipped with a Maxwell coil allowing for additional linear B0-field variations in z-direction. The effect of increasing strength of this superimposed gradient on image quality was observed using a standard single-shot EPI-based DWI sequence (sEPI), a zoomed single-shot EPI sequence (zEPI), a readout-segmented EPI sequence (rsEPI), and an sEPI sequence with integrated dynamic shimming (intEPI) on a 3-Tesla system. Additionally, ten volunteers were examined over the neck region using these techniques. Image quality was assessed by two radiologists. Scan durations were recorded.

Results: With increasing strength of the external gradient, marked distortions, signal loss, and failure of fat suppression were observed using sEPI, zEPI, and rsEPI. These artifacts were markedly reduced using intEPI. Significantly better in vivo image quality was also observed using intEPI compared with the other techniques. Scan time of intEPI was similar to sEPI and zEPI and shorter than rsEPI.

Conclusion: The use of integrated 2D shim and frequency adjustment for EPI-based DWI results in a significant improvement in image quality of the head/neck region at 3 Tesla. Combining integrated shimming with rsEPI or zEPI can be expected to provide additional improvements.

Keywords: Artifacts; Diffusion weighted; Echo-planar Imaging; MRI; Neck.

Publication types

  • Comparative Study

MeSH terms

  • Adipose Tissue / chemistry
  • Adult
  • Artifacts
  • Diffusion Magnetic Resonance Imaging / methods*
  • Echo-Planar Imaging / methods*
  • Female
  • Humans
  • Image Enhancement / methods
  • Image Interpretation, Computer-Assisted / methods
  • Magnetic Fields*
  • Male
  • Phantoms, Imaging
  • Reproducibility of Results
  • Signal-To-Noise Ratio
  • Tomography, X-Ray Computed
  • Water / chemistry

Substances

  • Water