Repeatability of swept-source optical coherence tomography retinal and choroidal thickness measurements in neovascular age-related macular degeneration

Br J Ophthalmol. 2017 May;101(5):603-608. doi: 10.1136/bjophthalmol-2016-308999. Epub 2016 Aug 4.

Abstract

Background: The aim was to determine the intrasession repeatability of swept-source optical coherence tomography (SS-OCT)-derived retinal and choroidal thickness measurements in eyes with neovascular age-related macular degeneration (nAMD).

Methods: A prospective study consisting of patients with active nAMD enrolled in the Distance of Choroid Study at Moorfields Eye Hospital, London. Patients underwent three 12×9 mm macular raster scans using the deep range imaging (DRI) OCT-1 SS-OCT (Topcon) device in a single imaging session. Retinal and choroidal thicknesses were calculated for the ETDRS macular subfields. Repeatability was calculated according to methods described by Bland and Altman.

Results: 39 eyes of 39 patients with nAMD were included with a mean (±SD) age of 73.9 (±7.2) years. The mean (±SD) retinal thickness of the central macular subfield was 225.7 μm (±12.4 μm). The repeatability this subfield, expressed as a percentage of the mean central macular subfield thickness, was 23.2%. The percentage repeatability of the other macular subfields ranged from 13.2% to 28.7%. The intrasession coefficient of repeatability of choroidal thickness of the central macular subfield was 57.2 μm with a mean choroidal thickness (±SD) of 181 μm (±15.8 μm).

Conclusions: This study suggests that a change >23.2% of retinal thickness and 57.2 μm choroidal thickness in the central macular subfield is required to distinguish true clinical change from measurement variability when using the DRI OCT-1 device to manage patients with nAMD.

Keywords: Imaging; Macula; Neovascularisation; Retina.

Publication types

  • Research Support, Non-U.S. Gov't

MeSH terms

  • Aged
  • Aged, 80 and over
  • Choroid / pathology*
  • Female
  • Humans
  • Male
  • Middle Aged
  • Prospective Studies
  • Reproducibility of Results
  • Retina / pathology*
  • Tomography, Optical Coherence / methods*
  • Wet Macular Degeneration / diagnostic imaging
  • Wet Macular Degeneration / pathology*