Post-Mortem Corneal Thickness Measurements with a Portable Optical Coherence Tomography System: a Reliability Study

Sci Rep. 2016 Jul 26:6:30428. doi: 10.1038/srep30428.

Abstract

The purpose of this study was to determine the repeatability and reproducibility of post-mortem central corneal thickness (CCT) measurements by using a real-time, portable optical coherence tomography (OCT) system on an animal model, and to prospectively evaluate the time-course of post-mortem changes in CCT. Forty-six ocular globes of sheep (Ovis aries) were analyzed with a portable spectral-domain OCT device by two operators at different postmortem intervals (PMIs) as follows: immediately (i.e. within 10 minutes), at the 30(th) minute, at the 1(st), 6(th), 12(th), 24(th) and 48(th) hour, and later (up to the 96(th) hour). The coefficient of repeatability ranged from 0.3% to 3.5%, and coefficients of reproducibility ranged from 0.2% to 3.7% in the central region of the cornea. The intraclass correlation coefficients were particularly high at different PMIs, thus confirming good measurement reliability with the portable OCT. The average CCT decreased immediately and then increased thereafter, with two peaks at 6 and 24 hours after death. Our results suggest that portable OCT is a reliable tool for monitoring CCT variations after death and may be useful in characterizing corneas before explantation, detecting quantitative variations during post-mortem corneal degeneration or assessing changes in CCT for forensic implications.

MeSH terms

  • Animals
  • Cornea / anatomy & histology
  • Corneal Pachymetry / methods*
  • Humans
  • Postmortem Changes*
  • Reproducibility of Results
  • Sheep
  • Tomography, Optical Coherence*