Simultaneous Determination of Hydrochlorothiazide and Losartan Potassium in Osmotic Pump Tablets by Microemulsion Liquid Chromatography

J Chromatogr Sci. 2016 Sep;54(8):1415-20. doi: 10.1093/chromsci/bmw101. Epub 2016 Jun 21.

Abstract

A rapid and efficient oil-in-water microemulsion liquid chromatographic (MELC) method has been optimized and validated for the determination of hydrochlorothiazide (HCT) and losartan potassium (LOP) in osmotic pump tablets. Samples were injected into a C18 (150 mm × 4.6 mm ID, 5 µm particle size) analytical column, which was maintained at 30°C. The most effective MELC system had a mobile phase consisting of 95% (v/v) of 3.0% (w/w) SDS, 6.0% (w/w) n-butanol, 0.8% (w/w) n-octane, 90.2% (w/w) water and 5% (v/v) acetonitrile (pH 5). The flow rate was 1.0 mL min(-1) and UV detection was performed at 265 nm. Linearity ranged from 2.5 to 12.5 µg mL(-1) for HCT and 10.0-60.0 µg mL(-1) for LOP (r > 0.999 for both drugs). The proposed method was rapid, precise (RSDs < 1.4%) and accurate (98.9% recovery for HCT and 101% recovery for LOP). It is applicable to simultaneous determination of HCT and LOP in osmotic pump tablets.

Publication types

  • Validation Study

MeSH terms

  • Chemistry Techniques, Analytical / instrumentation
  • Chemistry Techniques, Analytical / methods*
  • Chromatography, High Pressure Liquid
  • Chromatography, Liquid*
  • Emulsions / chemistry
  • Hydrochlorothiazide / analysis*
  • Losartan / analysis*
  • Reproducibility of Results

Substances

  • Emulsions
  • Hydrochlorothiazide
  • Losartan