In Vivo Response of Laser Processed Porous Titanium Implants for Load-Bearing Implants

Ann Biomed Eng. 2017 Jan;45(1):249-260. doi: 10.1007/s10439-016-1673-8. Epub 2016 Jun 15.

Abstract

Applications of porous metallic implants to enhance osseointegration of load-bearing implants are increasing. In this work, porous titanium implants, with 25 vol.% porosity, were manufactured using Laser Engineered Net Shaping (LENS™) to measure the influence of porosity towards bone tissue integration in vivo. Surfaces of the LENS™ processed porous Ti implants were further modified with TiO2 nanotubes to improve cytocompatibility of these implants. We hypothesized that interconnected porosity created via additive manufacturing will enhance bone tissue integration in vivo. To test our hypothesis, in vivo experiments using a distal femur model of male Sprague-Dawley rats were performed for a period of 4 and 10 weeks. In vivo samples were characterized via micro-computed tomography (CT), histological imaging, scanning electron microscopy, and mechanical push-out tests. Our results indicate that porosity played an important role to establish early stage osseointegration forming strong interfacial bonding between the porous implants and the surrounding tissue, with or without surface modification, compared to dense Ti implants used as a control.

Keywords: Additive manufacturing; LENS™; Osseointegration; Porous Ti; TiO2 Nanotubes.

Publication types

  • Research Support, N.I.H., Extramural

MeSH terms

  • Animals
  • Implants, Experimental*
  • Lasers*
  • Male
  • Materials Testing*
  • Nanotubes / chemistry*
  • Porosity
  • Rats
  • Rats, Sprague-Dawley
  • Titanium* / chemistry
  • Titanium* / pharmacology
  • Weight-Bearing
  • X-Ray Microtomography

Substances

  • titanium dioxide
  • Titanium