Inframammary Fold Reconstruction: A Biomechanical Analysis

Plast Reconstr Surg Glob Open. 2016 Mar 7;4(3):e634. doi: 10.1097/GOX.0000000000000568. eCollection 2016 Mar.

Abstract

Background: Inframammary fold reconstruction has scarcely been evaluated in literature. No biomechanical analyses have been performed comparing different reconstructive methods. This evaluation compares the gold-standard suture reconstruction with an intrarib anchor system (Micro BioComposite SutureTak, Arthrex).

Methods: Three analysis groups were compared including 8 Sawbone blocks, 22 embalmed cadaver, and 27 regular cadaver specimens (N = 57). Transient mechanical analysis was performed at 5 N/s using an Instron 5565 test frame.

Results: Ultimate load favored the anchor system (compared with the gold-standard suture) by a factor of 9.8 (P < 0.0001) for the regular cadaver group and a factor of 1.7 (P < 0.038) for the embalmed cadaver group. A similar statistically significant benefit was shown for stiffness and load at 2-mm displacement.

Conclusions: This analysis showed an anchor system to be the biomechanically superior fixation method in terms of ultimate load, fixation stiffness, and displacement at failure when compared with the gold-standard suture method in inframammary fold reconstruction. Because of superior stability in every aspect, an anchor system may be considered for inframammary fold reconstruction.