Impact of stent platform on longitudinal stent deformation: an in vivo frequency domain optical coherence tomography study

Cardiovasc Interv Ther. 2017 Jul;32(3):199-205. doi: 10.1007/s12928-016-0403-3. Epub 2016 May 25.

Abstract

Recently, longitudinal coronary stent deformation has been highlighted as a possible cause of drug-eluting stent failure. Although bench tests and in vivo studies have demonstrated difference in longitudinal stent strength among the stents with different platforms, its clinical impact is still unknown. Furthermore, it is unknown if modified stent platform favorably affect the incidence of stent deformation. The aim of this study was to investigate the longitudinal deformation of the everolimus-eluting stents (EES) with different stent platforms by using frequency domain optical coherence tomography (FD-OCT). Seventy-eight lesions treated with EES (Xience Prime: n = 26, Promus element: n = 29, Promus premier: n = 23) were studied. After successful stent implantation, FD-OCT was performed and stent length was measured using three-dimensional reconstruction of the images in vivo. Percent longitudinal stent shortening (%SS) was defined as the in vivo stent length divided by nominal stent length. Longitudinal stent deformation was defined as %SS > 10 %. Patients' and procedural characteristics were similar among 3 EESs. There was no difference in mean %SS between Xience Prime, Promus Element and Promus Premier (1.0 ± 5.8, 2.9 ± 6.7 and 0.8 ± 3.7 %, p = 0.322). Incidence of the longitudinal stent deformation was significantly higher in Promus Element than the other stents (0, 13.8 and 0 %, p = 0.028). Incidence of longitudinal stent deformation was different between EESs with different stent platforms. Stent material, stent design and/or stent delivery balloon may affect longitudinal stent deformation.

Keywords: Deformation; Drug-eluting stent fracture; Optical coherence tomography.

MeSH terms

  • Aged
  • Angina Pectoris / surgery
  • Blood Vessel Prosthesis Implantation / adverse effects
  • Blood Vessel Prosthesis Implantation / methods
  • Blood Vessel Prosthesis* / adverse effects
  • Drug-Eluting Stents* / adverse effects
  • Equipment Failure Analysis
  • Everolimus / therapeutic use
  • Female
  • Humans
  • Immunosuppressive Agents / therapeutic use
  • Longitudinal Studies
  • Male
  • Prosthesis Failure
  • Radiography, Interventional
  • Tomography, Optical Coherence

Substances

  • Immunosuppressive Agents
  • Everolimus