Effect of automated tube voltage selection, integrated circuit detector and advanced iterative reconstruction on radiation dose and image quality of 3rd generation dual-source aortic CT angiography: An intra-individual comparison

Eur J Radiol. 2016 May;85(5):972-8. doi: 10.1016/j.ejrad.2016.02.021. Epub 2016 Feb 24.

Abstract

Purpose: To compare, on an intra-individual basis, the effect of automated tube voltage selection (ATVS), integrated circuit detector and advanced iterative reconstruction on radiation dose and image quality of aortic CTA studies using 2nd and 3rd generation dual-source CT (DSCT).

Material and methods: We retrospectively evaluated 32 patients who had undergone CTA of the entire aorta with both 2nd generation DSCT at 120kV using filtered back projection (FBP) (protocol 1) and 3rd generation DSCT using ATVS, an integrated circuit detector and advanced iterative reconstruction (protocol 2). Contrast-to-noise ratio (CNR) was calculated. Image quality was subjectively evaluated using a five-point scale. Radiation dose parameters were recorded.

Results: All studies were considered of diagnostic image quality. CNR was significantly higher with protocol 2 (15.0±5.2 vs 11.0±4.2; p<.0001). Subjective image quality analysis revealed no significant differences for evaluation of attenuation (p=0.08501) but image noise was rated significantly lower with protocol 2 (p=0.0005). Mean tube voltage and effective dose were 94.7±14.1kV and 6.7±3.9mSv with protocol 2; 120±0kV and 11.5±5.2mSv with protocol 1 (p<0.0001, respectively).

Conclusion: Aortic CTA performed with 3rd generation DSCT, ATVS, integrated circuit detector, and advanced iterative reconstruction allow a substantial reduction of radiation exposure while improving image quality in comparison to 120kV imaging with FBP.

Keywords: Aorta; Automated tube voltage selection; CT angiography; Dual-source CT; Radiation dose reduction.

Publication types

  • Comparative Study
  • Evaluation Study

MeSH terms

  • Aged
  • Aorta / diagnostic imaging*
  • Aortic Aneurysm / diagnostic imaging
  • Aortic Dissection / diagnostic imaging
  • Aortography / methods
  • Aortography / standards
  • Computed Tomography Angiography / methods
  • Computed Tomography Angiography / standards*
  • Contrast Media
  • Female
  • Humans
  • Male
  • Middle Aged
  • Observer Variation
  • Radiation Dosage
  • Radiographic Image Interpretation, Computer-Assisted / methods
  • Retrospective Studies

Substances

  • Contrast Media