The validity of visual acuity assessment using mobile technology devices in the primary care setting

Aust Fam Physician. 2016 Apr;45(4):212-5.

Abstract

Background: The assessment of visual acuity is indicated in a number of clinical circumstances. It is commonly conducted through the use of a Snellen wall chart. Mobile technology developments and adoption rates by clinicians may potentially provide more convenient methods of assessing visual acuity. Limited data exist on the validity of these devices and applications.

Objective: The objective of this study was to evaluate the assessment of distance visual acuity using mobile technology devices against the commonly used 3-metre Snellen chart in a primary care setting.

Methods: A prospective quantitative comparative study was conducted at a regional medical practice. The visual acuity of 60 participants was assessed on a Snellen wall chart and two mobile technology devices (iPhone, iPad).

Results: Visual acuity intervals were converted to logarithm of minimum angle of resolution (logMAR) scores and subjected to intraclass correlation coefficient (ICC) assessment. The results show a high level of general agreement between testing modality (ICC 0.917 with a 95% confidence interval of 0.887-0.940).

Discussion: The high level of agreement of visual acuity results between the Snellen wall chart and both mobile technology devices suggests that clinicians can use this technology with confidence in the primary care setting.

Publication types

  • Comparative Study
  • Validation Study

MeSH terms

  • Adult
  • Aged
  • Aged, 80 and over
  • Female
  • Humans
  • Male
  • Middle Aged
  • Mobile Applications*
  • Primary Health Care / methods*
  • Prospective Studies
  • Smartphone*
  • Vision Tests / instrumentation*
  • Visual Acuity*
  • Young Adult