Cocaine and benzoylecgonine oral fluid on-site screening and confirmation

Drug Test Anal. 2016 Mar-Apr;8(3-4):296-303. doi: 10.1002/dta.1966. Epub 2016 Mar 9.

Abstract

Accurate on-site devices to screen for drug intake are critical for establishing whether an individual is driving under the influence of drugs (DUID); however, on-site oral fluid (OF) cocaine device performance is variable. We evaluated the performance of a newly developed benzoylecgonine (BE) test-strip for the Draeger® DrugTest 5000 device (20 µg/L cut-off) with equivalent cross reactivity for cocaine and BE. Ten cocaine users provided OF, collected with the Draeger cassette and Oral-Eze® and StatSure Saliva Sampler(TM) devices, up to 69 h following 25 mg intravenous cocaine administration. All screening results were confirmed by a validated two-dimensional-gas chromatography-mass spectrometry (2D-GC-MS) method for cocaine and/or BE. Cocaine test-strip median Tlast for screening only results was 6.5 h, and 6.5 h with Oral-Eze® and 4 h for StatSure OF confirmation for cocaine and/or BE at 1, 8, and 10 µg/L; sensitivity, specificity, and efficiency ranged from 85.5 to 100% and 83.3 to 100% for cocaine only confirmation at 8 and 10 µg/L. For the BE test-strip, median Tlast was 12.5 h for screening only and confirmation for cocaine and/or BE at all three cut-offs; sensitivity, specificity, and efficiency ranged from 85.5 to 97.5% and 78.4 to 97.4% with cocaine and/or BE confirmation at 8 and 10 µg/L cut-offs, respectively. The Draeger cocaine test-strip with cocaine only confirmation offers a useful option for monitoring the acute intoxication phase of DUID; additionally the BE test-strip with cocaine and/or BE confirmation increases the length of detection of cocaine intake for workplace drug testing, drug court, parole, pain management, drug treatment programs and both the acute cocaine intoxication and cocaine crash/fatigue phase of DUID. Copyright © 2016 John Wiley & Sons, Ltd.

Keywords: Draeger DrugTest 5000; benzoylecgonine; cocaine; on-site testing; oral fluid.

Publication types

  • Research Support, N.I.H., Intramural

MeSH terms

  • Adult
  • Cocaine / analogs & derivatives*
  • Cocaine / analysis*
  • Female
  • Gas Chromatography-Mass Spectrometry
  • Humans
  • Male
  • Middle Aged
  • Saliva / chemistry*
  • Sensitivity and Specificity
  • Substance Abuse Detection / instrumentation
  • Substance Abuse Detection / methods*

Substances

  • benzoylecgonine
  • Cocaine