[A Novel Evaluation Method for Displacement between Carbon Beam Axis and Positioning X-ray Axis Using an Imaging Plate]

Nihon Hoshasen Gijutsu Gakkai Zasshi. 2016 Feb;72(2):139-48. doi: 10.6009/jjrt.2016_JSRT_72.2.139.
[Article in Japanese]

Abstract

Purpose: We developed an evaluation method for easily calculating displacement directly between the carbon beam axis and positioning X-ray axis.

Methods: A verification image was acquired by irradiating an imaging plate with a carbon beam and X-ray. The X-ray passed through a lead plate inserted in the range compensator holder. The displacement was calculated on the verification image from the center of a wire irradiated with carbon using a multi leaf collimator (MLC) and a wire irradiated with X-ray also using MLC. The accuracy of the method was evaluated by moving the carbon beam axis, the X-ray axis, and the setup angle. The weekly changes of vertical and lateral beams in all rooms were also evaluated.

Results: The displacements of the carbon beam axis and the setup angle did not influence the calculation results, whereas the displacement of the X-ray axis did (R=0.999). The displacements including weekly changes were all less than 1.00 mm.

Conclusion: An evaluation method for calculating the displacement directly and simply between the carbon beam axis and positioning X-ray axis was developed and verified. The weekly changes of displacement between axes were evaluated to be acceptable at our facility.