Coverage analysis in a targeted amplicon-based next-generation sequencing panel for myeloid neoplasms

J Clin Pathol. 2016 Sep;69(9):801-4. doi: 10.1136/jclinpath-2015-203580. Epub 2016 Feb 19.

Abstract

Aims: PCR amplicon-based next-generation sequencing (NGS) panels are increasingly used for clinical diagnostic assays. Amplification bias is a well-known limitation of PCR amplicon-based approaches. We sought to characterise lower-performance amplicons in an off-the-shelf NGS panel (TruSight Myeloid Sequencing Panel) for myeloid neoplasms and attempted to patch the low read depth for one of the affected genes, CEBPA.

Methods: We performed targeted NGS of 158 acute myeloid leukaemia samples and analysed the amplicon read depths across 568 amplicons to identify lower-performance amplicons. We also correlated the amplicon read depths with the template GC content. Finally, we attempted to patch the low read depth for CEBPA using a parallel library preparation (Nextera XT) workflow.

Results: We identified 16 lower-performance amplicons affecting nine genes, including CEBPA. There was a slight negative correlation between the amplicon read depths and template GC content. Addition of the separate CEBPA library generated a minimum read depth per base across the CEBPA gene ranging from 268x to 758x across eight samples.

Conclusions: The identification of lower-performance amplicons will be informative to laboratories intending to use this panel. We have also demonstrated proof-of-concept that different libraries (TruSight Myeloid and Nextera XT) can be combined and sequenced on the same flow cell to generate additional reads for CEBPA.

Keywords: GENETICS; LEUKAEMIA; MOLECULAR PATHOLOGY.