Right Bundle Branch Block-Like Pattern During Uncomplicated Right Ventricular Pacing and the Effect of Pacing Site

Am J Cardiol. 2016 Mar 15;117(6):935-9. doi: 10.1016/j.amjcard.2015.12.028. Epub 2015 Dec 30.

Abstract

Right bundle branch block (RBBB) configuration is an unexpected finding during right ventricular (RV) pacing that raises the suspicion of inadvertent left ventricular lead positioning. The aim of this study was to evaluate the prevalence of paced RBBB pattern in relation to RV lead location. This is a secondary analysis of a prospective, multicenter study, which randomized implantable cardioverter defibrillator recipients to an apical versus midseptal defibrillator lead positioning. A 12-lead electrocardiogram was recorded during intrinsic rhythm and RV pacing. Paced RBBB-like pattern was defined as positive (>0.05 mV) net amplitude of QRS complex in leads V1 and/or V2. In total, 226 patients (65.6 ± 12.0 years, 20.8% women, 53.1% apical site) were included in the study. The prevalence of paced RBBB pattern in the total population was 15.5%. A significantly lower percentage of patients in the midseptal group demonstrated RBBB-type configuration during RV pacing compared with the apical group (1.9% vs 27.5%, p <0.001). Baseline RBBB, prolonged QRS duration during intrinsic rhythm, and reduced ejection fraction were not associated with increased likelihood of paced RBBB. In the subgroup of patients with RBBB type during pacing, 91.4% of patients had a paced QRS axis from -30° to -90°, whereas 100% of patients displayed a negative QRS vector at lead V3. In conclusion, RBBB configuration is encountered in a considerable percentage of device recipients during uncomplicated RV pacing. Midseptal lead positioning is associated with significantly lower likelihood of paced RBBB pattern compared with apical location.

Publication types

  • Comparative Study
  • Multicenter Study
  • Randomized Controlled Trial

MeSH terms

  • Aged
  • Bundle-Branch Block / etiology
  • Bundle-Branch Block / physiopathology*
  • Bundle-Branch Block / therapy*
  • Defibrillators, Implantable / adverse effects*
  • Diabetes Complications / physiopathology
  • Diabetes Complications / therapy
  • Electrocardiography
  • Female
  • Germany / epidemiology
  • Greece / epidemiology
  • Heart Conduction System / physiopathology*
  • Heart Ventricles*
  • Humans
  • Hypertension / complications
  • Male
  • Middle Aged
  • Myocardial Ischemia / therapy
  • Pacemaker, Artificial / adverse effects*
  • Prevalence
  • Prospective Studies
  • Research Design
  • Risk Factors
  • Treatment Outcome