Area densitometry using rotating Scheimpflug photography for posterior capsule opacification and surface light scattering analyses

J Cataract Refract Surg. 2015 Nov;41(11):2444-9. doi: 10.1016/j.jcrs.2015.05.038.

Abstract

Purpose: To compare area densitometry analysis using rotating Scheimpflug photography in quantifications of posterior capsule opacification (PCO) and surface light scattering with previous anterior-segment analyzer measurement.

Setting: Miyata Eye Hospital, Miyazaki, Japan.

Design: Prospective observational case series.

Methods: Scheimpflug images of eyes with foldable intraocular lenses (IOLs) were obtained using rotating and fixed Scheimpflug photography. Area densitometry on the posterior and anterior surfaces was conducted for PCO and surface light scattering analyses, respectively, with an identical area size. Correlation between two measurements was analyzed using linear regression.

Results: The study included 105 eyes of 74 patients who received IOLs 1 to 18 years (mean, 4.9 ± 4.5 years) postoperatively. In the PCO analysis on the posterior IOL surface, there was a significant correlation between the two measurements (P < .001, R(2) = 0.60). In the surface light scattering analysis, a significant and higher correlation was obtained (P < .001, R(2) = 0.91) until the fixed Scheimpflug photography exhibited saturation due to intensive scatterings.

Conclusion: Area densitometry combined with a rotating Scheimpflug photography was exchangeable to previously established densitometry measurement, and allowed successive evaluation in longer-term observations.

Publication types

  • Observational Study

MeSH terms

  • Aged
  • Aged, 80 and over
  • Capsule Opacification / diagnosis*
  • Densitometry / methods*
  • Female
  • Humans
  • Lens Implantation, Intraocular
  • Light
  • Male
  • Phacoemulsification
  • Photography / instrumentation*
  • Posterior Capsule of the Lens / pathology*
  • Prospective Studies
  • Scattering, Radiation*