Exposure dose reduction during lateral spine test with water filter

Technol Health Care. 2016 May 18;24(3):429-38. doi: 10.3233/THC-151120.

Abstract

To minimize exposure dose during lateral spine X-Ray testing and obtain optimal image for diagnosis a water filter was made to measure and evaluate dose distribution. When applying the water filter, as thickness increased exposure dose decreased. When applying 2 cm of water filter, clarity of contrast and boundary was found to be 4.5 ± 0.5 and resolution was found to be 2.00 ± 0.5 Lp/mm which was almost identical image quality compared to not applying water filter which showed clarity of contrast and boundary of 5.0 ± 0.0, and resolution of 2.50 ± 0.0 Lp/mm, while reducing exposure dose by 55%. This result is expected to have many uses as important basic data to predict exposure dose of patients and to minimize medical exposure dose through applying water filters during lateral spine X-Ray testing.

Keywords: Water filter; dose reduce; exposure dose; image quality.

MeSH terms

  • Humans
  • Paper*
  • Radiation Dosage*
  • Radiation Dosimeters
  • Radiation Injuries / prevention & control*
  • Spine / diagnostic imaging*
  • Tomography, X-Ray Computed / methods*