Calculation of crystalline lens power using a modification of the Bennett method

Biomed Opt Express. 2015 Oct 21;6(11):4501-15. doi: 10.1364/BOE.6.004501. eCollection 2015 Nov 1.

Abstract

We present a method for measuring lens power from extended depth OCT biometry, corneal topography, and refraction using an improvement on the Bennett method. A reduced eye model was used to derive a formula for lens power in terms of ocular distances, corneal power, and objective spherical equivalent refraction. An error analysis shows that the formula predicts relaxed lens power with a theoretical accuracy of ± 0.5 D for refractive error ranging from -10 D to + 10 D. The formula was used to calculate lens power in 16 eyes of 8 human subjects. Mean lens power was 24.3 D ± 1.7 D.

Keywords: (170.4500) Optical coherence tomography; (170.4580) Optical diagnostics for medicine; (330.7325) Visual optics, metrology; (330.7326) Visual optics, modeling.