Assessment of dose error due to nylon mesh of treatment couch

Phys Med. 2015 Dec;31(8):1080-1084. doi: 10.1016/j.ejmp.2015.08.013. Epub 2015 Oct 1.

Abstract

Purpose: This study aims at the assessment of dose error in patients undergoing radiotherapy due to treatment couch of Co-60 teletherapy unit.

Materials and methods: In this study beam attenuation due to treatment couch of Co-60 unit was measured in air for different gantry angles and field sizes. Polymethylmethacrylate (PMMA) phantom was used to estimate the effect of depth on attenuation. Impact of couch on surface dose was also evaluated.

Results: Beam attenuation due to couch was in the range of 0.5-28% for different gantry angles with standard field size of 10 × 10 cm(2) with optimum position of metallic cranks. Maximum attenuation (29%) was observed with smallest field size i.e. 5 × 5 cm(2). Beam attenuation has been found higher in phantom as compared to that in air However, no particular trend of attenuation has been noted with varying depth of phantom. A 6% increase in surface dose has also been observed due to couch insertion for normal beam incidence. Maximum error of 80% is also note-worthy for most unfavorable situation of irradiation at 180 degree through the metallic cranks.

Conclusion: It has been determined that ignoring the treatment couch and its accessories can result in dose error of 0.5-80%, depending on gantry angle, field size and position of couch accessories. Therefore, consideration of dose error due to couch during treatment planning is recommended.

Keywords: Attenuation; Dose error; Skin dose; Treatment couch.

MeSH terms

  • Nylons*
  • Radiation Dosage*
  • Radioisotope Teletherapy / instrumentation*
  • Radiotherapy Dosage
  • Radiotherapy Setup Errors*
  • Skin / radiation effects

Substances

  • Nylons