Left atrial appendage closure: First in man with the 4th generation WATCHMAN device

Catheter Cardiovasc Interv. 2016 Mar;87(4):787-94. doi: 10.1002/ccd.26238. Epub 2015 Sep 10.

Abstract

Objectives: To evaluate the feasibility and safety of the fourth generation WATCHMAN device.

Background: The WATCHMAN left atrial appendage (LAA) closure device has been shown to be non-inferior to conventional oral anticoagulation with warfarin for stroke prevention in patients with non-valvular atrial fibrillation. A new (fourth) generation of the WATCHMAN device was designed to facilitate easier delivery and improve safety.

Methods: We conducted a prospective, non-randomized study of LAA closure with use of the 4(th) generation WATCHMAN device in 36 patients with non-valvular atrial fibrillation. Follow-up was performed at 45 days, 6 months and 12 months following implantation.

Results: The mean age of the population was 73 ± 6 years and 67% were male. The mean CHADS2 and CHA2 DS2-VASc Score were 2.5 and 4.5, respectively. The device was implanted successfully in 34/36 (94%). No serious hospital complications related to the device or procedure occurred. During 12 month follow-up, three ischemic strokes (8%), one hemorrhagic stroke (3%), one TIA (3%) and three deaths (9%) occurred. Device-associated thrombus formation was detected in one patient (3%) during six-month follow-up and was treated successfully with low molecular weight heparin.

Conclusions: LAA closure with the 4(th) generation WATCHMAN device is feasible and safe. New technical features may contribute to a better performance of the occluder, particularly regarding the risk of pericardial effusions.

Keywords: atrial fibrillation; left atrial appendage closure; stroke prevention.

Publication types

  • Evaluation Study

MeSH terms

  • Aged
  • Aged, 80 and over
  • Atrial Appendage* / diagnostic imaging
  • Atrial Appendage* / physiopathology
  • Atrial Fibrillation / complications
  • Atrial Fibrillation / mortality
  • Atrial Fibrillation / physiopathology
  • Atrial Fibrillation / therapy*
  • Cardiac Catheterization / adverse effects
  • Cardiac Catheterization / instrumentation*
  • Cardiac Catheterization / mortality
  • Echocardiography, Transesophageal
  • Feasibility Studies
  • Female
  • Follow-Up Studies
  • Humans
  • Ischemic Attack, Transient / etiology
  • Ischemic Attack, Transient / mortality
  • Ischemic Attack, Transient / prevention & control*
  • Male
  • Middle Aged
  • Prospective Studies
  • Prosthesis Design*
  • Prosthesis Implantation / adverse effects
  • Prosthesis Implantation / instrumentation*
  • Prosthesis Implantation / mortality
  • Stroke / etiology
  • Stroke / mortality
  • Stroke / prevention & control*
  • Thrombosis / etiology
  • Time Factors
  • Treatment Outcome