Feasibility of CAIPIRINHA-Dixon-TWIST-VIBE for dynamic contrast-enhanced MRI of the prostate

Eur J Radiol. 2015 Nov;84(11):2110-6. doi: 10.1016/j.ejrad.2015.08.013. Epub 2015 Aug 22.

Abstract

Purpose: To evaluate the feasibility of a CAIPIRINHA-Dixon-TWIST (CDT)-VIBE sequence for improving image quality and temporal resolution in dynamic contrast-enhanced MRI (DCE-MRI) of the prostate.

Material and methods: 44 male patients (age 63.9 ± 8.9 years) with clinically suspected prostate cancer underwent DCE-MRI at a 3T MRI scanner (Magnetom Skyra, Siemens Healthcare, Erlangen, Germany) using a CDT-VIBE sequence (spatial resolution=3×1.2×1.2mm(3), temporal resolution=5s, total scan duration=4:10 min) with body-weight-adapted administration of contrast agent (Gadobutrol, Bayer Healthcare, Berlin, Germany). To investigate effects on image quality, the same sequence was acquired three times per patient during the late phase: 1. with the same protocol as in the arterial phase (VS5), 2. without view-sharing (no view-sharing, NVS) using a 2-fold CAIPIRINHA acceleration R=2 (temporal resolution=15s, NVS15) and 3. NVS using a 6-fold CAIPIRINHA acceleration R=6 (temporal resolution=5s, NVS5). SNR and CNR were evaluated with the subtraction method. Image quality of the three sequences (VS5, NVS15, NVS5) was subjectively assessed by 2 blinded radiologists using a 5-point Likert scale (5 being excellent). Perfusion profiles of visually normal prostate and of malignant lesions as characterized by Wash-In, Wash-Out, time-to-peak (TTP) and initial area under the curve (iAUC) from the original datasets (temporal resolution=5s) and from datasets with a downsampled temporal resolution (15s) were compared.

Results: In 20 of 44 included patients, potentially malignant lesions were identified in which 16 had histologically confirmed prostate cancer. SNR was highest for VS5 and NVS15, and lowest for NVS5 (p <.001). Concordantly, subjective image quality was comparable for VS5 and NVS15 (sum score 23.20 ± 1.03 vs 23.53 ± 1.34) and significantly lower for NVS5 (sum score 9.83 ± 2.32; p<.001). Perfusion parameters of macroscopically normal prostate tissue and suspect lesions differed significantly between original datasets and datasets with simulated lower temporal resolution, with the latter showing higher Wash-In (p=.002), lower Wash-Out (p=.003), higher Time-to-Peak (p<.001) and lower iAUC (p<.001).

Conclusion: CDT-VIBE can be readily exploited for DCE-MRI of the prostate preserving the diagnostic image quality while providing high temporal resolution for quantitative diagnostic assessment of enhancement curves in malignant lesions.

Keywords: CDT-VIBE; DCE-MRI; Dynamic imaging; Magnetic resonance imaging; Prostate; Prostate cancer.

MeSH terms

  • Aged
  • Analysis of Variance
  • Contrast Media*
  • Feasibility Studies
  • Humans
  • Image Enhancement / methods*
  • Image Interpretation, Computer-Assisted / methods*
  • Magnetic Resonance Imaging / methods*
  • Male
  • Middle Aged
  • Observer Variation
  • Organometallic Compounds
  • Prostate / pathology
  • Prostatic Neoplasms / pathology*
  • Reproducibility of Results

Substances

  • Contrast Media
  • Organometallic Compounds
  • gadobutrol