Automatic estimation and correction of anisotropic magnification distortion in electron microscopes

J Struct Biol. 2015 Nov;192(2):204-8. doi: 10.1016/j.jsb.2015.08.006. Epub 2015 Aug 13.

Abstract

We demonstrate a significant anisotropic magnification distortion, found on an FEI Titan Krios microscope and affecting magnifications commonly used for data acquisition on a Gatan K2 Summit detector. We describe a program (mag_distortion_estimate) to automatically estimate anisotropic magnification distortion from a set of images of a standard gold shadowed diffraction grating. We also describe a program (mag_distortion_correct) to correct for the estimated distortion in collected images. We demonstrate that the distortion present on the Titan Krios microscope limits the resolution of a set of rotavirus VP6 images to ∼7 Å, which increases to ∼3 Å following estimation and correction of the distortion. We also use a 70S ribosome sample to demonstrate that in addition to affecting resolution, magnification distortion can also interfere with the classification of heterogeneous data.

Keywords: Electron cryo-microscopy; Image correction; Magnification anisotropy; Resolution.

Publication types

  • Research Support, Non-U.S. Gov't

MeSH terms

  • Anisotropy*
  • Cryoelectron Microscopy / instrumentation
  • Cryoelectron Microscopy / methods*
  • Gold / analysis*
  • Image Enhancement / methods*