Improved in vivo performance of amperometric oxygen (PO2) sensing catheters via electrochemical nitric oxide generation/release

Anal Chem. 2015 Aug 18;87(16):8067-72. doi: 10.1021/acs.analchem.5b01590. Epub 2015 Aug 5.

Abstract

A novel electrochemically controlled release method for nitric oxide (NO) (based on electrochemical reduction of nitrite ions) is combined with an amperometric oxygen sensor within a dual lumen catheter configuration for the continuous in vivo sensing of the partial pressure of oxygen (PO2) in blood. The on-demand electrochemical NO generation/release method is shown to be fully compatible with amperometric PO2 sensing. The performance of the sensors is evaluated in rabbit veins and pig arteries for 7 and 21 h, respectively. Overall, the NO releasing sensors measure both venous and arterial PO2 values more accurately with an average deviation of -2 ± 11% and good correlation (R(2) = 0.97) with in vitro blood measurements, whereas the corresponding control sensors without NO release show an average deviation of -31 ± 28% and poor correlation (R(2) = 0.43) at time points >4 h after implantation in veins and >6 h in arteries. The NO releasing sensors induce less thrombus formation on the catheter surface in both veins and arteries (p < 0.05). This electrochemical NO generation/release method could offer a new and attractive means to improve the biocompatibility and performance of implantable chemical sensors.

Publication types

  • Research Support, N.I.H., Extramural

MeSH terms

  • Animals
  • Biosensing Techniques / methods*
  • Electrochemistry / trends
  • Monitoring, Physiologic / methods*
  • Nitric Oxide / blood
  • Nitric Oxide / chemistry*
  • Oxygen / analysis*
  • Rabbits
  • Swine

Substances

  • Nitric Oxide
  • Oxygen