Hydrazide and hydrazine reagents as reactive matrices for matrix-assisted laser desorption/ionization mass spectrometry to detect steroids with carbonyl groups

Eur J Mass Spectrom (Chichester). 2015;21(2):79-90. doi: 10.1255/ejms.1336.

Abstract

Fifty-one reagents (37 hydrazides and 14 hydrazines) were examined to determine if they could function as reactive matrices for the detection of steroids with carbonyl groups in matrix-assisted laser desorption/ionization mass spectrometry (MALDI-MS). Among these reagents, 3-hydroxy-2-naphthoic acid hydrazide reacted most efficiently with steroids with carbonyl groups, allowing detection of them at the 1pmol level. Results from accurate mass measurement with a JMS-S3000 Spiral-TOF suggested that the ion peaks corresponding to the protonated fragment of a hydrazone were observed, in addition to the protonated molecular peaks of the resulting hydrazones ([M(D)+H](+)) and their sodium adduct ([M(D)+Na](+)). 3-Hydroxy-2-naphthoic acid hydrazide also could detect cortisol (1.93ng) spiked in a human serum certified reference material, as well as cortisone and cortisol in a human saliva sample collected in the morning. Using a combination of 3-hydroxy-2-naphthoic acid hydrazide and α- cyano-4-hydroxycinammic acid (CHCA) (3-hydroxy-2-naphthoic acid hydrazide:CHCA=4:1, v/v), even cortisone and cortisol in a human saliva sample collected in the afternoon, when levels are typically low, could be detected effectively.