MR-based attenuation correction for cardiac FDG PET on a hybrid PET/MRI scanner: comparison with standard CT attenuation correction

Eur J Nucl Med Mol Imaging. 2015 Sep;42(10):1574-80. doi: 10.1007/s00259-015-3089-3. Epub 2015 Jun 20.

Abstract

Purpose: The aim of this study was to evaluate the feasibility of attenuation correction (AC) for cardiac (18)F-labelled fluorodeoxyglucose (FDG) positron emission tomography (PET) using MR-based attenuation maps.

Methods: We included 23 patients with no known cardiac history undergoing whole-body FDG PET/CT imaging for oncological indications on a PET/CT scanner using time-of-flight (TOF) and subsequent whole-body PET/MR imaging on an investigational hybrid PET/MRI scanner. Data sets from PET/MRI (with and without TOF) were reconstructed using MR AC and semi-quantitative segmental (20-segment model) myocardial tracer uptake (per cent of maximum) and compared to PET/CT which was reconstructed using CT AC and served as standard of reference.

Results: Excellent correlations were found for regional uptake values between PET/CT and PET/MRI with TOF (n = 460 segments in 23 patients; r = 0.913; p < 0.0001) with narrow Bland-Altman limits of agreement (-8.5 to +12.6 %). Correlation coefficients were slightly lower between PET/CT and PET/MRI without TOF (n = 460 segments in 23 patients; r = 0.851; p < 0.0001) with broader Bland-Altman limits of agreement (-12.5 to +15.0 %). PET/MRI with and without TOF showed minimal underestimation of tracer uptake (-2.08 and -1.29 %, respectively), compared to PET/CT.

Conclusion: Relative myocardial FDG uptake obtained from MR-based attenuation corrected FDG PET is highly comparable to standard CT-based attenuation corrected FDG PET, suggesting interchangeability of both AC techniques.

Publication types

  • Comparative Study
  • Evaluation Study

MeSH terms

  • Adult
  • Aged
  • Algorithms
  • Artifacts*
  • Cardiac Imaging Techniques / instrumentation
  • Cardiac Imaging Techniques / methods*
  • Feasibility Studies
  • Female
  • Fluorodeoxyglucose F18 / pharmacokinetics*
  • Humans
  • Image Enhancement / instrumentation
  • Image Enhancement / methods*
  • Magnetic Resonance Imaging / instrumentation
  • Magnetic Resonance Imaging / methods*
  • Male
  • Middle Aged
  • Multimodal Imaging / instrumentation
  • Multimodal Imaging / methods
  • Myocardium / metabolism
  • Positron-Emission Tomography / instrumentation
  • Positron-Emission Tomography / methods*
  • Radiopharmaceuticals
  • Reproducibility of Results
  • Sensitivity and Specificity
  • Tomography, X-Ray Computed / methods

Substances

  • Radiopharmaceuticals
  • Fluorodeoxyglucose F18