An Adaptive Thresholding Method for BTV Estimation Incorporating PET Reconstruction Parameters: A Multicenter Study of the Robustness and the Reliability

Comput Math Methods Med. 2015:2015:571473. doi: 10.1155/2015/571473. Epub 2015 May 19.

Abstract

Objective: The aim of this work was to assess robustness and reliability of an adaptive thresholding algorithm for the biological target volume estimation incorporating reconstruction parameters.

Method: In a multicenter study, a phantom with spheres of different diameters (6.5-57.4 mm) was filled with (18)F-FDG at different target-to-background ratios (TBR: 2.5-70) and scanned for different acquisition periods (2-5 min). Image reconstruction algorithms were used varying number of iterations and postreconstruction transaxial smoothing. Optimal thresholds (TS) for volume estimation were determined as percentage of the maximum intensity in the cross section area of the spheres. Multiple regression techniques were used to identify relevant predictors of TS.

Results: The goodness of the model fit was high (R(2): 0.74-0.92). TBR was the most significant predictor of TS. For all scanners, except the Gemini scanners, FWHM was an independent predictor of TS. Significant differences were observed between scanners of different models, but not between different scanners of the same model. The shrinkage on cross validation was small and indicative of excellent reliability of model estimation.

Conclusions: Incorporation of postreconstruction filtering FWHM in an adaptive thresholding algorithm for the BTV estimation allows obtaining a robust and reliable method to be applied to a variety of different scanners, without scanner-specific individual calibration.

Publication types

  • Multicenter Study
  • Research Support, Non-U.S. Gov't

MeSH terms

  • Algorithms
  • Computational Biology
  • Fluorodeoxyglucose F18
  • Head and Neck Neoplasms / diagnostic imaging
  • Humans
  • Image Processing, Computer-Assisted / methods
  • Linear Models
  • Models, Statistical
  • Phantoms, Imaging
  • Positron-Emission Tomography / statistics & numerical data*
  • Radiopharmaceuticals
  • Reproducibility of Results
  • Tomography, X-Ray Computed

Substances

  • Radiopharmaceuticals
  • Fluorodeoxyglucose F18