Whole-Body PET/MR Imaging: Quantitative Evaluation of a Novel Model-Based MR Attenuation Correction Method Including Bone

J Nucl Med. 2015 Jul;56(7):1061-6. doi: 10.2967/jnumed.115.156000. Epub 2015 May 29.

Abstract

In routine whole-body PET/MR hybrid imaging, attenuation correction (AC) is usually performed by segmentation methods based on a Dixon MR sequence providing up to 4 different tissue classes. Because of the lack of bone information with the Dixon-based MR sequence, bone is currently considered as soft tissue. Thus, the aim of this study was to evaluate a novel model-based AC method that considers bone in whole-body PET/MR imaging.

Methods: The new method ("Model") is based on a regular 4-compartment segmentation from a Dixon sequence ("Dixon"). Bone information is added using a model-based bone segmentation algorithm, which includes a set of prealigned MR image and bone mask pairs for each major body bone individually. Model was quantitatively evaluated on 20 patients who underwent whole-body PET/MR imaging. As a standard of reference, CT-based μ-maps were generated for each patient individually by nonrigid registration to the MR images based on PET/CT data. This step allowed for a quantitative comparison of all μ-maps based on a single PET emission raw dataset of the PET/MR system. Volumes of interest were drawn on normal tissue, soft-tissue lesions, and bone lesions; standardized uptake values were quantitatively compared.

Results: In soft-tissue regions with background uptake, the average bias of SUVs in background volumes of interest was 2.4% ± 2.5% and 2.7% ± 2.7% for Dixon and Model, respectively, compared with CT-based AC. For bony tissue, the -25.5% ± 7.9% underestimation observed with Dixon was reduced to -4.9% ± 6.7% with Model. In bone lesions, the average underestimation was -7.4% ± 5.3% and -2.9% ± 5.8% for Dixon and Model, respectively. For soft-tissue lesions, the biases were 5.1% ± 5.1% for Dixon and 5.2% ± 5.2% for Model.

Conclusion: The novel MR-based AC method for whole-body PET/MR imaging, combining Dixon-based soft-tissue segmentation and model-based bone estimation, improves PET quantification in whole-body hybrid PET/MR imaging, especially in bony tissue and nearby soft tissue.

Keywords: MR-based attenuation correction; PET/MR hybrid imaging; attenuation correction of bone; model-based attenuation correction.

Publication types

  • Research Support, N.I.H., Extramural
  • Research Support, Non-U.S. Gov't

MeSH terms

  • Adult
  • Aged
  • Aged, 80 and over
  • Algorithms
  • Bone and Bones / diagnostic imaging*
  • Bone and Bones / pathology
  • Female
  • Fluorodeoxyglucose F18 / chemistry
  • Humans
  • Image Processing, Computer-Assisted
  • Magnetic Resonance Imaging*
  • Male
  • Middle Aged
  • Multimodal Imaging*
  • Positron-Emission Tomography*
  • Radiopharmaceuticals / chemistry
  • Tomography, X-Ray Computed
  • Whole Body Imaging*

Substances

  • Radiopharmaceuticals
  • Fluorodeoxyglucose F18