Simultaneous monitoring of Staphylococcus aureus growth in a multi-parametric microfluidic platform using microscopy and impedance spectroscopy

Bioelectrochemistry. 2015 Oct:105:56-64. doi: 10.1016/j.bioelechem.2015.05.006. Epub 2015 May 13.

Abstract

We describe the design, construction, and characterization of a scalable microfluidic platform that allows continuous monitoring of biofilm proliferation under shear stress conditions. Compared to other previous end-point assay studies, our platform offers the advantages of integration into multiple environments allowing simultaneous optical microscopy and impedance spectroscopy measurements. In this work we report a multi-parametric sensor that can monitor the growth and activity of a biofilm. This was possible by combining two interdigitated microelectrodes (IDuEs), and punctual electrodes to measure dissolved oxygen, K+, Na+ and pH. The IDuE has been optimized to permit sensitive and reliable impedance monitoring of Staphylococcus aureus V329 growth with two- and four-electrode measurements. We distinguished structural and morphological changes on intact cellular specimens using four-electrode data modeling. We also detected antibiotic mediated effects using impedance. Results were confirmed by scanning electrode microscopy and fluorescence microscopy after live/dead cell staining. The bacitracin mediated effects detected with impedance prove that the approach described can be used for guiding the development of novel anti-biofilm agents to better address bacterial infection.

Keywords: Bacitracin; Biofilm; Impedance spectroscopy; Interdigitated microelectrodes; Microfluidic device.

Publication types

  • Research Support, Non-U.S. Gov't

MeSH terms

  • Biofilms
  • Electric Impedance
  • Microelectrodes
  • Microfluidics*
  • Spectrum Analysis / methods*
  • Staphylococcus aureus / growth & development*