Automated CD34+ cell isolation of peripheral blood stem cell apheresis product

Cytotherapy. 2015 Oct;17(10):1465-71. doi: 10.1016/j.jcyt.2015.04.005. Epub 2015 May 14.

Abstract

Background aims: Immunomagnetic enrichment of CD34+ hematopoietic "stem" cells (HSCs) using paramagnetic nanobead coupled CD34 antibody and immunomagnetic extraction with the CliniMACS plus system is the standard approach to generating T-cell-depleted stem cell grafts. Their clinical beneficence in selected indications is established. Even though CD34+ selected grafts are typically given in the context of a severely immunosuppressive conditioning with anti-thymocyte globulin or similar, the degree of T-cell depletion appears to affect clinical outcomes and thus in addition to CD34 cell recovery, the degree of T-cell depletion critically describes process quality. An automatic immunomagnetic cell processing system, CliniMACS Prodigy, including a protocol for fully automatic CD34+ cell selection from apheresis products, was recently developed. We performed a formal process validation to support submission of the protocol for CE release, a prerequisite for clinical use of Prodigy CD34+ products.

Methods: Granulocyte-colony stimulating factor-mobilized healthy-donor apheresis products were subjected to CD34+ cell selection using Prodigy with clinical reagents and consumables and advanced beta versions of the CD34 selection software. Target and non-target cells were enumerated using sensitive flow cytometry platforms.

Results: Nine successful clinical-scale CD34+ cell selections were performed. Beyond setup, no operator intervention was required. Prodigy recovered 74 ± 13% of target cells with a viability of 99.9 ± 0.05%. Per 5 × 10E6 CD34+ cells, which we consider a per-kilogram dose of HSCs, products contained 17 ± 3 × 10E3 T cells and 78 ± 22 × 10E3 B cells.

Conclusions: The process for CD34 selection with Prodigy is robust and labor-saving but not time-saving. Compared with clinical CD34+ selected products concurrently generated with the predecessor technology, product properties, importantly including CD34+ cell recovery and T-cell contents, were not significantly different. The automatic system is suitable for routine clinical application.

Keywords: CD34; CliniMACS; Prodigy; allogeneic; automation; cell therapy; clean room; good manufacturing practice; haplo-identical; immunomagnetic; naked haplo; stem cell transplantation.

Publication types

  • Research Support, Non-U.S. Gov't

MeSH terms

  • Antigens, CD34 / immunology*
  • Antilymphocyte Serum / immunology
  • Automation, Laboratory
  • B-Lymphocytes / immunology
  • Blood Component Removal / methods*
  • Cell Separation / methods*
  • Cells, Cultured
  • Flow Cytometry
  • Granulocyte Colony-Stimulating Factor / immunology
  • Hematopoietic Stem Cell Transplantation / methods
  • Hematopoietic Stem Cells / cytology*
  • Hematopoietic Stem Cells / immunology
  • Humans
  • Immunomagnetic Separation / methods*
  • Lymphocyte Depletion / methods
  • T-Lymphocytes / immunology

Substances

  • Antigens, CD34
  • Antilymphocyte Serum
  • Granulocyte Colony-Stimulating Factor