Coronary calcium scores are systematically underestimated at a large chest size: A multivendor phantom study

J Cardiovasc Comput Tomogr. 2015 Sep-Oct;9(5):415-21. doi: 10.1016/j.jcct.2015.03.010. Epub 2015 Apr 2.

Abstract

Objective: To evaluate the effect of chest size on coronary calcium score (CCS) as assessed with new-generation CT systems from 4 major vendors.

Methods: An anthropomorphic, small-sized (300 × 200 mm) chest phantom containing 100 small calcifications (diameters, 0.5-2.0 mm) was evaluated with and without an extension ring on state-of-the-art CT systems from 4 vendors. The extension ring was used to mimic a patient with a large chest size (400 × 300 mm). Image acquisition was repeated 5 times with small translations and/or rotations. Routine clinical acquisition and reconstruction protocols for small and large patients were used. CCS was quantified as Agatston and mass scores with vendor software.

Results: The small-sized phantom resulted in median (interquartiles) Agatston scores of 10 (9-35), 136 (123-146), 34 (30-37), and 87 (85-89) for Philips, GE, Siemens, and Toshiba, respectively. Mass scores were 4 mg (3-9 mg), 23 mg (21-27 mg), 8 mg (8-9 mg), and 20 mg (20-20 mg), respectively. Adding the extension ring resulted in reduced Agatston scores for all vendors (17%-48%) and mass scores for 2 vendors (11%-49%). Median Agatston scores decreased to 9 (5-10), 79 (60-80), 27 (24-32), and 45 (29-53) units, and median mass scores remained similar for Philips at 4 mg (4-6 mg) and Siemens at 8 mg (7-8 mg) and decreased for the other vendors to 13 mg (11-14 mg) and 10 mg (8-13 mg), respectively.

Conclusion: This multivendor phantom study showed that CCS can be underestimated up to 50% (49%-66%) for Agatston scores and 49% (36%-59%) for mass scores at a larger chest size, which may be relevant for women and large patients. However, CCS underestimation by chest size differs considerably by vendor.

Keywords: Agatston score; Chest size; Computed tomography; Coronary calcium score.

Publication types

  • Comparative Study
  • Research Support, Non-U.S. Gov't

MeSH terms

  • Coronary Angiography / instrumentation*
  • Coronary Artery Disease / diagnostic imaging*
  • Coronary Vessels / diagnostic imaging*
  • Equipment Design
  • Humans
  • Phantoms, Imaging*
  • Predictive Value of Tests
  • Radiographic Image Interpretation, Computer-Assisted
  • Reproducibility of Results
  • Severity of Illness Index
  • Software
  • Thorax / anatomy & histology*
  • Tomography Scanners, X-Ray Computed*
  • Tomography, X-Ray Computed / instrumentation*
  • Vascular Calcification / diagnostic imaging*