A fully-automated, six-plex single molecule immunoassay for measuring cytokines in blood

J Immunol Methods. 2015 Sep:424:20-7. doi: 10.1016/j.jim.2015.04.017. Epub 2015 May 7.

Abstract

We report a system and assay for performing fully-automated measurement of 6 proteins simultaneously with single molecule sensitivity. The system combines handling of samples, reagents, and consumables, with a module for imaging single molecule arrays (Simoa) to enable immunoassays that have high sensitivity (~fg/mL), are multiplexed, and are fully-automated. A 6-plex cytokine Simoa assay for IL-6, TNF-α, GM-CSF, IL-10, IL-1β, and IL-1α was developed on the system. The assays had limits of detection in the range 0.01-0.03pg/mL, and the average imprecision (CV) of the Simoa signal was 4.2%. This assay was used to measure the concentrations of these cytokines in the plasma of patients with Crohn's Disease (CD), before and after treatment with anti-TNF-α antibody drugs, and in the serum of Type 1 diabetics. Concentrations of TNF-α and IL-6 in the CD samples determined using the fully-automated, multiplex Simoa assay had good correlation with the manual, single-plex assays previously reported. Drug treatment caused reductions in the mean concentration of all 6 cytokines in the plasma of CD patients. The concentrations of 4 cytokines were significantly higher in diabetics compared to healthy controls. The system could enable the widespread, multiplexed measurement of protein biomarkers with low abundance.

Keywords: Automation; Digital ELISA; Multiplexed immunoassays; Simoa.

Publication types

  • Research Support, N.I.H., Extramural

MeSH terms

  • Automation*
  • Crohn Disease / blood
  • Cytokines / blood*
  • Diabetes Mellitus / blood
  • Enzyme-Linked Immunosorbent Assay / instrumentation
  • Enzyme-Linked Immunosorbent Assay / methods
  • Humans
  • Immunoassay / instrumentation
  • Immunoassay / methods*
  • Reproducibility of Results
  • Workflow

Substances

  • Cytokines