Development and validation of a DESI-HRMS/MS method for the fast profiling of esomeprazole and its metabolites in rat plasma: a pharmacokinetic study

Drug Test Anal. 2016 Feb;8(2):208-13. doi: 10.1002/dta.1805. Epub 2015 Apr 28.

Abstract

The advances in pharmaceutical development and drug discovery impose the availability of reliable high-throughput screening methods for the rapid evaluation of drug metabolism and pharmacokinetic (PK) in biological samples. Here, a desorption electrospray mass spectrometry (DESI-MS) method has been developed and validated for the PK profiling of esomeprazole and its metabolites (5-hydroxyomeprazole and omeprazole sulfone) in rat plasma. Rats were treated with an esomeprazole solution (2.5 mg/mL) for endovenous administration and the analyte levels were profiled over 2 h after liquid-liquid extraction from plasma. MS and tandem mass spectrometry (MS/MS) experiments were performed by using a DESI-LTQ-Orbitrap XL instrument and an on-spot fixed time analysis on PMMA surfaces. Validation was performed for the esomeprazole. The DESI-MS/MS method exhibited for the esomepazole excellent sensitivity (limit of detection (LOD)=60 ng/mL), linearity (0.2-20 µg/mL concentration range; y=23848(±361)X, n=15; r(2) =0.987) and precision (RSD<9%) by using an internal standard method. The PK results were discussed in terms of Area Under the Curve, Cmax and Tmax . Data reliability was demonstrated by comparison with a liquid chromatography-tandem mass spectrometry method (p>0.05). The data achieved demonstrated that the DESI-MS method is suitable for sensitive and fast profiling of a drug and its metabolites at the therapeutic concentration levels.

Keywords: DESI-HRMS; esomeprazole; pharmacokinetic.

MeSH terms

  • Animals
  • Area Under Curve
  • Biotransformation
  • Chromatography, High Pressure Liquid
  • Esomeprazole / blood*
  • Esomeprazole / pharmacokinetics*
  • Limit of Detection
  • Liquid-Liquid Extraction
  • Male
  • Omeprazole / analogs & derivatives
  • Proton Pump Inhibitors / blood*
  • Proton Pump Inhibitors / pharmacokinetics*
  • Rats
  • Rats, Wistar
  • Reference Standards
  • Reproducibility of Results
  • Spectrometry, Mass, Electrospray Ionization
  • Tandem Mass Spectrometry

Substances

  • Proton Pump Inhibitors
  • omeprazole sulfone
  • Omeprazole
  • Esomeprazole