Estimation of effective dose during hystrosalpingography procedures in certain hospitals in Sudan

Appl Radiat Isot. 2015 Jun:100:2-6. doi: 10.1016/j.apradiso.2015.02.009. Epub 2015 Feb 13.

Abstract

The aims of this study were to measure the patients' entrance surface air kerma doses (ESAK), effective doses and to compare practices between different hospitals in Sudan. ESAK were measured for patient using calibrated thermo luminance dosimeters (TLDs, GR200A). Effective doses were estimated using National radiological Protection Board (NRPB) software. This study was conducted in five radiological departments: Two Teaching Hospitals (A and D), two private hospitals (B and C) and one University Hospital (E). The mean ESAK was 20.1mGy, 28.9mGy, 13.6mGy, 17.5mGy, 35.7mGy for hospitals A, B, C, D, and E, respectively. The mean effective dose was 2.4mSv, 3.5mSv, 1.6mSv, 2.1mSv and 4.3mSv in the same order. The study showed wide variations in the ESDs with three of the hospitals having values above the internationally reported values.

Keywords: Effective dose; HSG; Patient dosimetry; Radiation risks; TLD.