Purpose: To evaluate and compare the mechanical properties of anterior capsule opening performed with femtosecond laser capsulotomy at different energy settings in ex vivo porcine anterior lens capsule specimens.

Methods: Twenty-five fresh porcine eyes per group were included in the study. Femtosecond laser capsulotomy was performed with three different pulse energy levels: 2 µJ (low energy group), 5 µJ (intermediate energy group), and 10 µJ (high energy group). The capsule openings were stretched with universal testing equipment until they ruptured. The morphologic profile of the cut capsule edges was evaluated using scanning electron microscopy.

Results: The high energy group had significantly lower rupture force (108 ± 14 mN) compared to the intermediate energy group (118 ± 10 mN) (P < .05) and low energy group (119 ± 11 mN) (P < .05), but the difference between the intermediate energy and low energy groups was not significant (P = .9479). The high energy group had significantly lower circumference stretching ratio (144% ± 3%) compared to the intermediate energy group (148% ± 3%) (P < .05) and low energy group (148% ± 3%) (P < .05), but the difference between the intermediate energy group and low energy group was not significant (P = .9985). Scanning electron microscopy images showed that the edge was only serrated with low and intermediate energy, but additional signs of collagen melting and denaturation were observed at high energy.

Conclusions: Anterior capsule openings created at a high energy level were slightly weaker and less extensible than those created at low or intermediate levels, possibly due to the increased thermal effect of photo-disruption.

Publication types

  • Comparative Study
  • Evaluation Study

MeSH terms

  • Animals
  • Anterior Capsule of the Lens / physiology*
  • Anterior Capsule of the Lens / surgery
  • Anterior Capsule of the Lens / ultrastructure
  • Biomechanical Phenomena
  • Elasticity / physiology*
  • Microscopy, Electron, Scanning
  • Posterior Capsulotomy / methods*
  • Swine