Image monitoring of pharmaceutical blending processes and the determination of an end point by using a portable near-infrared imaging device based on a polychromator-type near-infrared spectrometer with a high-speed and high-resolution photo diode array detector

Molecules. 2015 Mar 3;20(3):4007-19. doi: 10.3390/molecules20034007.

Abstract

In the present study we have developed a new version (ND-NIRs) of a polychromator-type near-infrared (NIR) spectrometer with a high-resolution photo diode array detector, which we built before (D-NIRs). The new version has four 5 W halogen lamps compared with the three lamps for the older version. The new version also has a condenser lens with a shorter focal point length. The increase in the number of the lamps and the shortening of the focal point of the condenser lens realize high signal-to-noise ratio and high-speed NIR imaging measurement. By using the ND-NIRs we carried out the in-line monitoring of pharmaceutical blending and determined an end point of the blending process. Moreover, to determinate a more accurate end point, a NIR image of the blending sample was acquired by means of a portable NIR imaging device based on ND-NIRs. The imaging result has demonstrated that the mixing time of 8 min is enough for homogeneous mixing. In this way the present study has demonstrated that ND-NIRs and the imaging system based on a ND-NIRs hold considerable promise for process analysis.

Publication types

  • Research Support, Non-U.S. Gov't

MeSH terms

  • Chemistry, Pharmaceutical / methods*
  • Halogens / chemistry
  • Signal-To-Noise Ratio
  • Spectroscopy, Near-Infrared / methods
  • Technology, Pharmaceutical / methods*

Substances

  • Halogens