Proton-induced x-ray fluorescence CT imaging

Med Phys. 2015 Feb;42(2):900-7. doi: 10.1118/1.4906169.

Abstract

Purpose: To demonstrate the feasibility of proton-induced x-ray fluorescence CT (pXFCT) imaging of gold in a small animal sized object by means of experiments and Monte Carlo (MC) simulations.

Methods: First, proton-induced gold x-ray fluorescence (pXRF) was measured as a function of gold concentration. Vials of 2.2 cm in diameter filled with 0%-5% Au solutions were irradiated with a 220 MeV proton beam and x-ray fluorescence induced by the interaction of protons, and Au was detected with a 3 × 3 mm(2) CdTe detector placed at 90° with respect to the incident proton beam at a distance of 45 cm from the vials. Second, a 7-cm diameter water phantom containing three 2.2-diameter vials with 3%-5% Au solutions was imaged with a 7-mm FWHM 220 MeV proton beam in a first generation CT scanning geometry. X-rays scattered perpendicular to the incident proton beam were acquired with the CdTe detector placed at 45 cm from the phantom positioned on a translation/rotation stage. Twenty one translational steps spaced by 3 mm at each of 36 projection angles spaced by 10° were acquired, and pXFCT images of the phantom were reconstructed with filtered back projection. A simplified geometry of the experimental data acquisition setup was modeled with the MC TOPAS code, and simulation results were compared to the experimental data.

Results: A linear relationship between gold pXRF and gold concentration was observed in both experimental and MC simulation data (R(2) > 0.99). All Au vials were apparent in the experimental and simulated pXFCT images. Specifically, the 3% Au vial was detectable in the experimental [contrast-to-noise ratio (CNR) = 5.8] and simulated (CNR = 11.5) pXFCT image. Due to fluorescence x-ray attenuation in the higher concentration vials, the 4% and 5% Au contrast were underestimated by 10% and 15%, respectively, in both the experimental and simulated pXFCT images.

Conclusions: Proton-induced x-ray fluorescence CT imaging of 3%-5% gold solutions in a small animal sized water phantom has been demonstrated for the first time by means of experiments and MC simulations.

Publication types

  • Research Support, N.I.H., Extramural

MeSH terms

  • Animals
  • Image Processing, Computer-Assisted
  • Monte Carlo Method
  • Optical Imaging / methods*
  • Phantoms, Imaging
  • Protons*
  • Tomography, X-Ray Computed / methods*
  • Water

Substances

  • Protons
  • Water