The analysis of linear and monomethylalkanes in exhaled breath samples by GC×GC-FID and GC-MS/MS

J Chromatogr B Analyt Technol Biomed Life Sci. 2015 Jan 26:978-979:62-9. doi: 10.1016/j.jchromb.2014.11.026. Epub 2014 Dec 3.

Abstract

A new arrangement of the INCAT (inside needle capillary adsorption trap) device with Carbopack X and Carboxen 1000 as sorbent materials was applied for sampling, preconcentration and injection of C6C19n-alkanes and their monomethyl analogs in exhaled breath samples. For the analysis both GC-MS/MS and GC×GC-FID techniques were used. Identification of the analytes was based on standards, measured retention indices and selective SRM transitions of the individual isomers. The GC-MS/MS detection limits were in the range from 2.1 pg for n-tetradecane to 86 pg for 5-methyloctadecane. The GC×GC-FID detection limits ranged from 19 pg for n-dodecane to 110 pg for 3-methyloctane.

Keywords: Exhaled breath; INCAT; Monomethylalkanes; Needle trap; n-Alkanes.

Publication types

  • Research Support, Non-U.S. Gov't

MeSH terms

  • Alkanes / analysis*
  • Alkanes / chemistry
  • Breath Tests / methods*
  • Gas Chromatography-Mass Spectrometry / methods*
  • Humans
  • Limit of Detection
  • Reproducibility of Results
  • Tandem Mass Spectrometry / methods