The aim of this study was to investigate the effect of ageing and fading on PADC detector response, as a function of the storage time and temperature. Several groups of CR-39 detectors provided by Radosys, Ltd. were exposed at the reference radon chamber of the Istituto Nazionale di Metrologia delle Radiazioni Ionizzanti, at the ENEA centre in Casaccia. The results indicate that low-temperature storage inhibits the effect of both ageing and fading. Finally, the overall reduction in CR-39 sensitivity due to the combined ageing/fading effect was estimated. In particular, the sensitivity of the detectors continuously exposed in air at room temperature over 6 and 3 months was reduced, respectively, by 7.5 and 4 %.
© The Author 2014. Published by Oxford University Press. All rights reserved. For Permissions, please email: journals.permissions@oup.com.