Trans-cranial focused ultrasound without hair shaving: feasibility study in an ex vivo cadaver model

J Ther Ultrasound. 2014 Jan 2:1:24. doi: 10.1186/2050-5736-1-24. eCollection 2013.

Abstract

In preparing a patient for a trans-cranial magnetic resonance (MR)-guided focused ultrasound procedure, current practice is to shave the patient's head on treatment day. Here we present an initial attempt to evaluate the feasibility of trans-cranial focused ultrasound in an unshaved, ex vivo human head model. A human skull filled with tissue-mimicking phantom and covered with a wig made of human hair was sonicated using 220- and 710-kHz head transducers to evaluate the feasibility of acoustic energy transfer. Heating at the focal point was measured by MR proton resonance shift thermometry. Results showed that the hair had a negligible effect on focal spot thermal rise at 220 kHz and a 17% drop in temperature elevation when using 710 kHz.

Keywords: Brain; Hair; MR thermometry; tcMRgFUS.