Endotracheal tubes for critically ill patients: an in vivo analysis of associated tracheal injury, mucociliary clearance, and sealing efficacy

Chest. 2015 May;147(5):1327-1335. doi: 10.1378/chest.14-1438.

Abstract

Background: Improvements in the design of the endotracheal tube (ETT) have been achieved in recent years. We evaluated tracheal injury associated with ETTs with novel high-volume low-pressure (HVLP) cuffs and subglottic secretions aspiration (SSA) and the effects on mucociliary clearance (MCC).

Methods: Twenty-nine pigs were intubated with ETTs comprising cylindrical or tapered cuffs and made of polyvinylchloride (PVC) or polyurethane. In specific ETTs, SSA was performed every 2 h. Following 76 h of mechanical ventilation, pigs were weaned and extubated. Images of the tracheal wall were recorded before intubation, at extubation, and 24 and 96 h thereafter through a fluorescence bronchoscope. We calculated the red-to-green intensity ratio (R/G), an index of tracheal injury, and the green-plus-blue (G+B) intensity, an index of normalcy, of the most injured tracheal regions. MCC was assessed through fluoroscopic tracking of radiopaque markers. After 96 h from extubation, pigs were killed, and a pathologist scored injury.

Results: Cylindrical cuffs presented a smaller increase in R/G vs tapered cuffs (P = .011). Additionally, cuffs made of polyurethane produced a minor increase in R/G (P = .012) and less G+B intensity decline (P = .022) vs PVC cuffs. Particularly, a cuff made of polyurethane and with a smaller outer diameter outperformed all cuffs. SSA-related histologic injury ranged from cilia loss to subepithelial inflammation. MCC was 0.9 ± 1.8 and 0.4 ± 0.9 mm/min for polyurethane and PVC cuffs, respectively (P < .001).

Conclusions: HVLP cuffs and SSA produce tracheal injury, and the recovery is incomplete up to 96 h following extubation. Small, cylindrical-shaped cuffs made of polyurethane cause less injury. MCC decline is reduced with polyurethane cuffs.

Publication types

  • Evaluation Study
  • Research Support, Non-U.S. Gov't

MeSH terms

  • Animals
  • Critical Illness / therapy*
  • Intubation, Intratracheal / adverse effects*
  • Intubation, Intratracheal / instrumentation*
  • Mucociliary Clearance
  • Swine
  • Trachea / injuries*
  • Trachea / physiology*
  • Treatment Outcome