RNA external standards, although important to ensure equivalence across many microarray platforms, have yet to be fully implemented in the research community. In this article, a set of unique RNA external standards (or RNA standards) and probe pairs that were added to total RNA in the samples before amplification and labeling are described. Concentration-response curves of RNA external standards were used across multiple commercial DNA microarray platforms and/or quantitative real-time polymerase chain reaction (RT-PCR) and next-generation sequencing to identify problematic assays and potential sources of variation in the analytical process. A variety of standards can be added in a range of concentrations spanning high and low abundances, thereby enabling the evaluation of assay performance across the expected range of concentrations found in a clinical sample. Using this approach, we show that we are able to confirm the dynamic range and the limit of detection for each DNA microarray platform, RT-PCR protocol, and next-generation sequencer. In addition, the combination of a series of standards and their probes was investigated on each platform, demonstrating that multiplatform calibration and validation is possible.
Keywords: Cross-platform; DNA microarray; Dynamic range; External RNA standards; Multiplatform calibration.