Force-controlled dynamic wear testing of total ankle replacements

Acta Biomater. 2015 Jan:12:332-340. doi: 10.1016/j.actbio.2014.10.036. Epub 2014 Oct 31.

Abstract

Currently, our knowledge of wear performance in total ankle replacements is limited. The aim of this study is to develop a scenario for force-controlled testing and wear testing of total ankle replacements. A force-controlled wear test was developed: based on cadaver measurements, the passive stabilization (ligaments and soft tissue) of the ankle joint was characterized and a restraint model for ankle stabilization was developed. Kinematics and kinetics acting at the replaced ankle joint were defined based on literature data and gait analysis. Afterwards, force-controlled wear testing was carried out on a mobile, three-component, total ankle replacement design. Wear was assessed gravimetrically and wear particles were analyzed. Wear testing resulted in a mean wear rate of 18.2±1.4mm(3)/10(6) cycles. Wear particles showed a mean size of 0.23μm with an aspect ratio of 1.61±0.96 and a roundness of 0.62±0.14. Wear testing of total ankle replacement shows that a relevant wear mass is generated with wear particles in a biologically relevant size range. The developed wear test provides a basis for future wear testing of total ankle replacements.

Keywords: Ankle arthroplasty; Ankle replacement; Wear; Wear particles; Wear simulation.

Publication types

  • Research Support, Non-U.S. Gov't

MeSH terms

  • Arthroplasty, Replacement, Ankle*
  • Biomechanical Phenomena
  • Humans
  • Materials Testing*