Early structural valve deterioration of Mitroflow aortic bioprosthesis: mode, incidence, and impact on outcome in a large cohort of patients

Circulation. 2014 Dec 2;130(23):2012-20. doi: 10.1161/CIRCULATIONAHA.114.010400. Epub 2014 Oct 29.

Abstract

Background: Structural valve deterioration (SVD) is a major flaw of bioprostheses. Early SVD has been suspected in the last models of Mitroflow bioprosthesis. We sought to assess the incidence, mode, and impact of SVD on outcome in a large series of Mitroflow aortic valve replacement.

Methods and results: Six hundred seventeen consecutive patients (aged 76.1±6.3 years) underwent aortic valve replacement with a Mitroflow prosthesis (models 12A/LX) between 2002 and 2007. By echocardiography, 39 patients developed early SVD (1.66% per patient-year), with stenosis as the main mode (n=36). Mean delay to SVD was only 3.8±1.4 years, and 5-year SVD-free survival was 91.6% (95% confidence interval [CI], 88.7-94.7) for the whole cohort and 79.8% (95% CI, 71.2-89.4) and 94.0% (95% CI, 90.3-97.8) for 19- and 21-mm sizes, respectively. Among the 39 patients with SVD, 13 patients (33%) had an accelerated SVD once the mean gradient exceeded 30 mm Hg. Valve-related death was 46.2% in this SVD subgroup. Five-year overall survival was 69.6% (95% CI, 65.7-73.9). In multivariable analysis, SVD was the strongest correlate of overall mortality (hazard ratio=7.7; 95% CI, 4.4-13.6).

Conclusions: Early SVD is frequent in Mitroflow bioprosthesis (models 12A/LX), especially for small sizes (19 and 21 mm), and reduces overall survival. An unpredictable accelerated pattern of SVD constitutes a life-threatening condition. In view of the large number of Mitroflow valves implanted worldwide, one can expect an epidemic of SVD and valve-related deaths, which represents a major public health issue, especially in the elderly. Hence, a close follow-up with yearly echocardiography after Mitroflow implantation is advisable. An urgent reoperation should be discussed in patients with severe SVD even though they are still asymptomatic.

Keywords: aortic stenosis; bioprosthesis; cardiac valves; echocardiography; survival analysis.

Publication types

  • Research Support, Non-U.S. Gov't

MeSH terms

  • Aged
  • Aged, 80 and over
  • Animals
  • Aortic Valve Stenosis / mortality*
  • Aortic Valve Stenosis / surgery*
  • Bioprosthesis / adverse effects
  • Bioprosthesis / statistics & numerical data*
  • Cattle
  • Female
  • Follow-Up Studies
  • Heart Failure / mortality
  • Heart Valve Prosthesis Implantation / adverse effects
  • Heart Valve Prosthesis Implantation / methods*
  • Heart Valve Prosthesis Implantation / mortality*
  • Humans
  • Incidence
  • Intensive Care Units / statistics & numerical data
  • Kaplan-Meier Estimate
  • Length of Stay / statistics & numerical data
  • Male
  • Neoplasms / mortality
  • Proportional Hazards Models
  • Reoperation / mortality
  • Shock, Cardiogenic / etiology
  • Shock, Cardiogenic / mortality