Study of a liquid plug-flow thermal cycling technique using a temperature gradient-based actuator

Sensors (Basel). 2014 Oct 27;14(11):20235-44. doi: 10.3390/s141120235.

Abstract

Easy-to-use thermal cycling for performing rapid and small-volume DNA amplification on a single chip has attracted great interest in the area of rapid field detection of biological agents. For this purpose, as a more practical alternative to conventional continuous flow thermal cycling, liquid plug-flow thermal cycling utilizes a thermal gradient generated in a serpentine rectangular flow microchannel as an actuator. The transit time and flow speed of the plug flow varied drastically in each temperature zone due to the difference in the tension at the interface between temperature gradients. According to thermal distribution analyses in microfluidics, the plug flow allowed for a slow heating process, but a fast cooling process. The thermal cycle of the microfluid was consistent with the recommended temperature gradient for PCR. Indeed, amplification efficiency of the plug flow was superior to continuous flow PCR, and provided an impressive improvement over previously-reported flow microchannel thermal cycling techniques.

Publication types

  • Research Support, Non-U.S. Gov't

MeSH terms

  • Algorithms
  • Equipment Design
  • Equipment Failure Analysis
  • Feedback
  • Heating / instrumentation*
  • Microfluidic Analytical Techniques / instrumentation*
  • Oscillometry / instrumentation*
  • Polymerase Chain Reaction / instrumentation*
  • Thermography / instrumentation*
  • Transducers*