Clinical impact of defibrillation testing at the time of implantable cardioverter-defibrillator insertion

Cardiol J. 2015;22(3):253-9. doi: 10.5603/CJ.a2014.0062.

Abstract

Background: Ventricular fibrillation is routinely induced during implantable cardioverter-defibrillator insertion to assess defibrillator performance, but this strategy is experiencing a progressive decline. We aimed to assess the efficacy of defibrillator therapies and long-term outcome in a cohort of patients that underwent defibrillator implantation with and without defibrillation testing.

Methods: Retrospective observational series of consecutive patients undergoing initial defibrillator insertion or generator replacement. We registered spontaneous ventricular arrhythmias incidence and therapy efficacy, and mortality.

Results: A total of 545 patients underwent defibrillator implantation (111 with and 434 without defibrillation testing). After 19 (range 9-31) months of follow-up, the death rate per observation year (4% vs. 4%; p = 0.91) and the rate of patients with defibrillator-treated ventricular arrhythmic events per observation year (with test: 10% vs. without test: 12%; p = 0.46) were similar. The generalized estimating equations-adjusted first shock probability of success in patients with test (95%; CI 88-100%) vs. without test (98%; CI 96-100%; p = 0.42) and the proportion of successful antitachycardia therapies (with test: 87% vs. without test: 80%; p = 0.35) were similar between groups. There was no difference in the annualized rate of failed first shock per patient and per shocked patient between groups (5% vs. 4%; p = 0.94).

Conclusions: In this observational study, that included an unselected population of patients with a defibrillator, no difference was found in overall mortality, first shock efficacy and rate of failed shocks regardless of whether defibrillation testing was performed or not.

Publication types

  • Observational Study

MeSH terms

  • Aged
  • Death, Sudden, Cardiac / prevention & control*
  • Defibrillators, Implantable*
  • Electric Countershock / adverse effects
  • Electric Countershock / instrumentation*
  • Electric Countershock / mortality
  • Electrophysiologic Techniques, Cardiac*
  • Female
  • Humans
  • Intraoperative Care / methods*
  • Male
  • Middle Aged
  • Predictive Value of Tests
  • Prosthesis Design
  • Prosthesis Failure
  • Prosthesis Implantation / instrumentation*
  • Retrospective Studies
  • Tachycardia, Ventricular / diagnosis
  • Tachycardia, Ventricular / physiopathology
  • Tachycardia, Ventricular / therapy*
  • Time Factors
  • Treatment Outcome
  • Ventricular Fibrillation / diagnosis
  • Ventricular Fibrillation / mortality
  • Ventricular Fibrillation / physiopathology
  • Ventricular Fibrillation / therapy*